Hello Ryan, with regards to the DiVa collection issue, we also observe this problem on our LSRII system. I have added two screenshots from measuring the same 8 peak bead sample, the problem is clearly recognizable when lots of plots are opened (the "busy" worksheet vs. the "light" worksheet). In our hands, the problem always occurs when the CPU usage of the HP xw 4300 workstation goes up to 100%. Therefore, as a workaround, we watch the CPU usage via the XP task manager and make sure that it is well below 100% by avoiding to open too many workspace items. Kind regards, Steffen Dr. Steffen Walter Director T-Cell Analytics Direct: +49 (0) 7071 565 125 40 *** Please find the contact details at the bottom of this mail *** -----Ursprüngliche Nachricht----- Von: Ryan Duggan [mailto:rduggan@bsd.uchicago.edu] Gesendet: Mittwoch, 30. April 2008 17:09 An: Cytometry Mailing List Betreff: DiVa data collection issue We've noticed a strange, but substantial quirk on our DiVa systems that we're wondering if others could corroborate. The symptom presented itself as a large discrepancy in the number of Threshold events (# of events recognized by the instrument) versus the number of events Processed (# of events that actually get processed and saved in your file). More specifically, the more events you collected (Threshold) the fewer events were actually saved (Processed). Typically, the reason this occurs is because of coincidence aborts. Since there is no dead time on DiVa instruments, the abort rate is usually only about 1%, assuming you have your window extension set properly. But, what we were seeing is that the processed event count 'plus' the abort count, which should be nearly equal to the threshold count, was way off, as much as 50% of the threshold count. After messing around for a while, the bottleneck ended up being the number of 'objects' being displayed on the worksheet. I'm using the term objects to refer to either a dot within a dot plot, a region, or simply a plot itself. The more of any of those object that the graphics card must process, the fewer the events it can process, therefore leading to a lot of data loss. It should follow then that if you display fewer objects, the data loss will decrease, and that's exactly what happened. We did one or two of the following to fix the problem: 1. Close all but a couple of dot plots in the worksheet to monitor integrity of data. 2. Change the "Events to Display" menu to a small number, e.g. 5000 events. 3. Do both. As soon as this was done, Processed events was right back to "normal" levels. This turns out to be a huge impact on overall yield for sorting as well because if it doesn't get processed, it doesn't get sorted. I suspect this is the reason why we've typically seen better yield on our MoFlo than we do on our Aria. I already knew that the sub par graphics card on these HPs caused the displaying of plots to lag after you collected your large files, but i didn't realize it was a bottleneck for data collection. Has anyone gone ahead and installed a high powered, really fast, graphics card in there standard HP to improve this issue? Can you send me info on what one(s) you've used? I'm using an HP XW4400s, and have noticed this on an Aria and LSRII. Do you think it's something other than the graphics processor? Thanks, Ryan ----------------------------------------------- Ryan Duggan Technical Director Flow Cytometry Facility University of Chicago Ph: 773.702.9212 or .5582 Web: http://ucflow.uchicago.edu Main Facility 910 E. 58th, Room 037 Chicago, IL 60637 ----------------------------------------------- This email is intended only for the use of the individual or entity to which it is addressed and may contain information that is privileged and confidential. If the reader of this email message is not the intended recipient, you are hereby notified that any dissemination, distribution, or copying of this communication is prohibited. If you have received this email in error, please notify the sender and destroy/delete all copies of the transmittal. Thank you. -- immatics biotechnologies GmbH Paul-Ehrlich-Str. 15 72076 Tuebingen, Germany Phone +49-7071-565125-0 Fax +49-7071-565125-99 www.immatics.com Commercial Register: Stuttgart HRB 38 2151 Managing Directors: Dr. Harpreet Singh, Dr. Niels Emmerich, Dr. Juergen Frisch This message may contain confidential and/or privileged information. If you are not the addressee or authorized to receive this for the addressee, you must not use, copy, disclose or take any action based on this message or any information herein. If you have received this message in error, please advise the sender immediately by reply e-mail and delete this message. Thank you for your cooperation. This attachment - 'Tube_001 HI 5o10 light sheet.JPG' - 627.06 KBytes - can be viewed at http://www.cyto.purdue.edu/MD-parts/062f46cc6b64a85dd5460485e8971de386347d7c.JPG This attachment - 'Tube_002 HI 5o10 busy sheet.JPG' - 773.55 KBytes - can be viewed at http://www.cyto.purdue.edu/MD-parts/3df3018a022d2cf089f1cb2631b1f36c1317d2af.JPGReceived on Mon May 5 12:58:00 2008
This archive was generated by hypermail 2.1.8 : Wed Jan 31 2007 - 03:12:00 EST